Some components in the N-Light™ tests are thermolabile. Therefore, performance and shelf life are assured only when the test is stored at 2-8 °C. The test can be exposed to higher temperatures for a short period of time as required for transport and handling. NEMIS has validated the anticipated worst-case storage conditions and confirms that the test performance will remain within the specifications for the times and temperatures listed below:
| Below 0°C | 2-8°C | 25°C | 30°C | 37°C | Above 37°C |
|---|---|---|---|---|---|
| Not allowed | Up to the expiry date | Up to 10 days | Up to 5 days | Up to 2 days | Not allowed |