What causes false positive (FP) or false negative (FN) results?

What causes false positive (FP) or false negative (FN) results?
Cause FP FN
Inappropriate storage conditions (see question 53)
Shorter incubation time
Incubate at a lower temperature than 37 °C
Incubate at a higher temperature than 37 °C
Measurement of the chemiluminescent signal without having incubated the sample according to the N-Light™ L.monocytogenes test Instructions For Use (IFU)
Measu Immediate or delayed measurement of the sample after tablet release contrary to the specifications in the N-Light™ L.monocytogenes test Instructions For Use (IFU)

Any deviation from the N-Light™ L.monocytogenes test IFU can lead to FP or FN results.
The growth of certain Bacilli, S.aureus, and L.ivanovii could theoretically cause positive signals as they produce the same enzyme PI-PLC which cleaves the AquaSpark™ molecule and leads to the expression of chemiluminescent light. However, the proprietary NEMIS enrichment broth contains both antibiotics and bacteriophages that specifically eliminate these microorganisms.