How important is it to stick to the 10 sec time between tablet release and luminometer reading stated in the IFU?

How important is it to stick to the 10 sec time between tablet release and luminometer reading stated in the IFU?

The reading should be started in 8 to 12 sec after table release to achieve minimal test to test variability. Shorter or longer times between tablet release and reading will lead to higher variability. Users should never wait for more than 20 sec before starting the read.